Unified Design Method Design of chromatographic separation processes

From dimensionless parameters to equipment

A feasible region on an operating parameter plane gives an operating point such as \((s_{\mathrm{cut}}, s^{\mathrm{FF}})\) or \((s_\mathrm{II}, S_\mathrm{II})\). The laboratory needs millilitres and minutes: how much to inject, when to cut, how fast to pump and when to switch. The conversion needs only the column volume, its porosity and one flow rate or switch time, and it is the same algebra for every process.

Symbols and “ideal conditions”: Notation and definitions. The Design converter does these calculations for you. Back to the Theory overview.

What the conversion needs

Every \(s\)-parameter is a volume of liquid divided by \(V_\mathrm{s}\), so volumes follow from \(V_\mathrm{s}\) alone and times from volumes and flow rates. Under ideal conditions the outlet profile on the \(s\) axis depends only on the isotherms and the feed, not on \(\varepsilon\), \(t_0\) or the flow rate. The purities of an operating point are therefore the same at any flow rate; the flow sets only the time scale. In practice the flow rate is limited by pressure drop and column efficiency, which the ideal model does not contain.

The definitions, solved for volumes and times

Unified Design parameters and the physical quantities they fix (Sainio 2016, Eqs. 5 to 7, 10, 12 and 42; flow rate \(Q\) during the event, \(V_\mathrm{s} = (1-\varepsilon)V_\mathrm{col}\)).
EventParameterVolume of liquidTime
FB event (feeding, eluting, a fraction)\(s_{\mathrm{FB}}\)\(Q\,\Delta t = V_\mathrm{s}\, s_{\mathrm{FB}}\)\(\Delta t = s_{\mathrm{FB}}\, t_0 F\)
Fresh feed\(s^{\mathrm{FF}}\)\(V^{\mathrm{FF}} = V_\mathrm{s}\, s^{\mathrm{FF}}\)\(\Delta t^{\mathrm{FF}} = s^{\mathrm{FF}}\, t_0 F\)
Cut event\(s_{\mathrm{cut}}\)\(Q\,t_\mathrm{cut} = Q\,\Delta t^{\mathrm{F}} + \varepsilon V_\mathrm{col} + V_\mathrm{s}\, s_{\mathrm{cut}}\)\(t_\mathrm{cut} = \Delta t^{\mathrm{F}} + t_0\,(1 + F s_{\mathrm{cut}})\)
Cut, batch elution\(S_{\mathrm{cut}} = s_{\mathrm{cut}} + s^{\mathrm{FF}}\)\(Q\,t_\mathrm{cut} = \varepsilon V_\mathrm{col}\,(1 + F S_{\mathrm{cut}})\)\(t_\mathrm{cut} = t_0\,(1 + F S_{\mathrm{cut}})\)
SMB event in zone \(j\)\(s_j\)\(Q_j\,\Delta t_\mathrm{switch} = \varepsilon V_\mathrm{col} + V_\mathrm{s}\, s_j\)\(\Delta t_\mathrm{switch} = t_{0,j}\,(1 + F s_j)\)

Times are measured from the start of the column feed; \(t_{0,j} = \varepsilon V_\mathrm{col}/Q_j\) is the dead time of one column at the flow rate of zone \(j\). In batch elution the column feed is the fresh feed, \(\Delta t^{\mathrm{F}} = \Delta t^{\mathrm{FF}}\).

Batch elution

The operating point \((s_{\mathrm{cut}}, s^{\mathrm{FF}})\) fixes the injection volume and the cut. With linear isotherms the bands also follow directly: component \(i\) leaves between \(s_{\mathrm{cut}} = H_i - s^{\mathrm{FF}}\) and \(H_i\), that is between \(t = t_0(1 + FH_i)\) and \(\Delta t^{\mathrm{FF}} + t_0(1 + FH_i)\). With stacked injections, the next injection starts so that its first band leaves just as the last band of the previous one has left; the cycle is \(t_\mathrm{cycle} = s_\mathrm{cycle}\, t_0 F\), with \(s_\mathrm{cycle}\) the width of the outlet profile on the \(s\) axis (Siitonen and Sainio, 2011, Eq. 34). The eluent used per cycle is \(V_\mathrm{s}(s_\mathrm{cycle} - s^{\mathrm{FF}})\).

Worked example

Linear isotherms with \(H_1 = 1.5\) and \(H_2 = 2.5\) (the system of the simple example), and the operating point \(s_{\mathrm{cut}} = 1.6\), \(s^{\mathrm{FF}} = 0.8\), inside the triangle \(1.5 \le s_{\mathrm{cut}} \le 2.5 - s^{\mathrm{FF}}\). The column has \(V_\mathrm{col} = 50\) mL and \(\varepsilon = 0.6\), so \(F = 2/3\), \(V_\mathrm{s} = 20\) mL and the liquid volume is 30 mL. At \(Q = 4\) mL/min the dead time is \(t_0 = 7.5\) min.

Batch elution of the worked example, checked against the exact solution of the ideal model.
QuantityFormulaValue
Injection volume\(V_\mathrm{s}\, s^{\mathrm{FF}}\)16 mL
Injection time\(s^{\mathrm{FF}}\, t_0 F\)4 min
Cut time\(t_0\,(1 + F S_{\mathrm{cut}})\), \(S_{\mathrm{cut}} = 2.4\)19.5 min
Component 1 leaves\(t_0(1 + FH_1)\) to \(\Delta t^{\mathrm{FF}} + t_0(1 + FH_1)\)15 to 19 min
Component 2 leaves\(t_0(1 + FH_2)\) to \(\Delta t^{\mathrm{FF}} + t_0(1 + FH_2)\)20 to 24 min
Cycle, stacked injections\(s_\mathrm{cycle}\, t_0 F\), \(s_\mathrm{cycle} = 1.8\)9 min
Eluent per cycle\(V_\mathrm{s}(s_\mathrm{cycle} - s^{\mathrm{FF}})\)20 mL

Product 1 is collected from 15 to 19.5 min and product 2 from 19.5 to 24 min; between 19 and 20 min nothing leaves, so a cut anywhere in that minute gives two pure products. That minute is the width of the feasible interval, \(2.5 - s^{\mathrm{FF}} - 1.5 = 0.2\), times \(t_0 F = 5\) min.

Batch elution of the equipment example over four cyclesInjections of 16 mL lasting 4 min every 9 min; component 1 leaves from 15 to 19 min, component 2 from 20 to 24 min; the cut is at 19.5 min; the next injection's first band starts exactly when the previous last band ends.Batch elution with stacked injections (times in minutes)inletoutlet16 mLcut 19.5product 1product 2t0369121518212427303336

Figure 1. The worked batch example in minutes over several cycles: injections of 16 mL (4 min) every 9 min, component 1 (blue) and component 2 (red) at the outlet, and the cut at 19.5 min (dashed). The first band of each injection leaves exactly when the last band of the previous one has left.

Four-zone SMB

A four-zone SMB needs four \(s\)-parameters, one per zone. Zones II and III come from the operating point: \(s_\mathrm{II}\), and \(s_\mathrm{III} = S_\mathrm{II} = s_\mathrm{II} + s^{\mathrm{FF}}\). Zones I and IV regenerate the adsorbent and the eluent; with linear isotherms this needs \(s_\mathrm{I} \ge H_2\) and \(s_\mathrm{IV} \le H_1\), and a margin is usual. Solving the definition of \(s_{\mathrm{SMB}}\) for the flow rate gives the zone flows, and the balances at the four nodes give the external streams (Sainio 2016, Eq. 12):

\[ Q_j = \frac{\varepsilon V_\mathrm{col} + V_\mathrm{s}\, s_j}{\Delta t_\mathrm{switch}}, \qquad j = \mathrm{I}, \ldots, \mathrm{IV}, \]
\[ Q^{\mathrm{FF}} = Q_\mathrm{III} - Q_\mathrm{II} = \frac{V_\mathrm{s}\, s^{\mathrm{FF}}}{\Delta t_\mathrm{switch}}, \quad Q^{\mathrm{EL}} = Q_\mathrm{I} - Q_\mathrm{IV}, \quad Q^{\mathrm{E}} = Q_\mathrm{I} - Q_\mathrm{II}, \quad Q^{\mathrm{R}} = Q_\mathrm{III} - Q_\mathrm{IV}. \]

The four zone flow rates, the four external streams and the switch time are fixed by the four \(s\)-values and the four node balances, except for one free choice. Usually it is the switch time, or the flow rate of zone I, the highest one, set to what the pump and the pressure drop allow. Each \(s_j\) refers to one column, so the formulas hold for any number of columns per zone. With \(N_\mathrm{SMB}\) SMB events per cycle (for example the Japan Organo process), \(S = s + s^{\mathrm{FF}}/N_\mathrm{SMB}\) and the fresh feed per cycle is \(V_\mathrm{s} N_\mathrm{SMB}(S - s)\) (Sainio 2016, Eqs. 9 and 59).

Worked example

The same point transferred to a four-zone SMB with four of the same columns, one per zone: \(s_\mathrm{II} = 1.6\), \(s_\mathrm{III} = 2.4\), and the margins \(s_\mathrm{I} = 3.0\) and \(s_\mathrm{IV} = 1.2\). With a switch time of 5 min:

Four-zone SMB of the worked example, \(\Delta t_\mathrm{switch} = 5\) min (computed in scripts/equipment.py).
Zone\(s_j\)\(Q_j\) (mL/min)NodeExternal stream (mL/min)
I3.018.0eluent in (before I)\(Q^{\mathrm{EL}} = 7.2\)
II1.612.4extract out (I/II)\(Q^{\mathrm{E}} = 5.6\)
III2.415.6feed in (II/III)\(Q^{\mathrm{FF}} = 3.2\)
IV1.210.8raffinate out (III/IV)\(Q^{\mathrm{R}} = 4.8\)

The balance closes: \(7.2 + 3.2 = 5.6 + 4.8\) mL/min. The fresh feed per switch is \(3.2 \times 5 = 16\) mL, the same as the batch injection on one column of the same size: same \(s^{\mathrm{FF}}\), same \(s_\mathrm{II} = s_{\mathrm{cut}}\), and under ideal conditions the same purities. If instead the pump limits zone I to 20 mL/min, the switch time becomes \((30 + 20 \times 3.0)/20 = 4.5\) min and the other zones run at 13.8, 17.3 and 12.0 mL/min; the \(s\)-values, and with them the purities, do not change.

Four-zone SMB of the example (mL/min, switch every 5 min)Four columns, one per zone, connected in a loop. Eluent enters before zone I, extract leaves between zones I and II, feed enters between zones II and III, raffinate leaves between zones III and IV, and the liquid from zone IV is recycled to zone I. After every switch time the columns move one position against the flow. Zone flow rates in mL/min: I 18, II 12.4, III 15.6, IV 10.8; external: eluent, extract, feed, raffinate 7.2, 5.6, 3.2, 4.8.Four-zone SMB of the example (mL/min, switch every 5 min)zone IsI = 3.0QI = 18.0zone IIsII = 1.6QII = 12.4zone IIIsIII = 2.4QIII = 15.6zone IVsIV = 1.2QIV = 10.8recycle (zone IV to zone I)eluent 7.2extract 5.6feed 3.2raffinate 4.8shift afterΔtswitch

Figure 2. The SMB of the worked example: four columns, one per zone, with the zone flow rates, the external streams and the recycle from zone IV to zone I (all in mL/min). Every 5 min the columns move one position against the flow (the ports move with it).

The design converter

The Design converter does these conversions for batch elution and the four-zone SMB, in both directions. Enter the column volume and porosity and the operating point (for batch elution also the flow rate or the dead time, for the SMB the switch time or one flow rate), and it returns the injection time and volume, the cut time, the zone and external flow rates and the switch time; or enter the settings and it returns the UD parameters. It uses the formulas above.

What the conversion does not tell you

The conversion is exact, but the operating point is only as good as the model behind it. Under ideal conditions the boundaries of the region are sharp, and a point on a boundary has no margin at all. Real columns have finite efficiency; the regions shrink, most at high purity and near the tip, and the flow rate now matters through the column efficiency and the pressure drop. Choose the operating point inside the feasible region with a margin, and check it with one experiment. Limits of the ideal picture discusses how large the margin should be.

Three equations printed differently in the papers

The formulas on this page follow from the definitions of the UD parameters. Three equations in the papers differ from them and are used here, and in the design converter and the examples, in the corrected form:

Further reading

  1. Sainio, T., Unified Design of chromatographic processes with timed events: Separation of ternary mixtures, Chem. Eng. Sci. 152 (2016) 547–567.
  2. Sainio, T., Unified design of chromatographic separation processes: Partial separations and waste fractions, Chem. Eng. Sci. 234 (2021) 116457. Open access.
  3. Siitonen, J., Sainio, T., Unified design of chromatographic separation processes, Chem. Eng. Sci. 122 (2015) 436–451.
  4. Siitonen, J., Sainio, T., Explicit equations for the height and position of the first component shock for binary mixtures with competitive Langmuir isotherms under ideal conditions, J. Chromatogr. A 1218 (2011) 6379–6387.
  5. Storti, G., Mazzotti, M., Morbidelli, M., Carrà, S., Robust design of binary countercurrent adsorption separation processes, AIChE J. 39 (1993) 471–492.

Disclaimer: No liability is assumed for correctness or completeness of data generated with the tools available on this website.