Design converter
An operating point is useful in the laboratory only as pump settings: a flow rate, an injection time and a cut time for batch elution, or four zone flows and a switch time for a simulated moving bed. The Unified Design parameters are dimensionless (they count the liquid pumped per volume of stationary phase), so the step to your column is pure arithmetic. This page does it in both directions, for batch elution and for the classical four-zone SMB. It is also the bridge for transferring a design: under ideal conditions the same \( s_\mathrm{cut} \) and \( s^\mathrm{FF} \) give the same purities in batch elution and in an SMB with \( s_\mathrm{II} = s_\mathrm{cut} \) and \( s_\mathrm{III} = s_\mathrm{cut} + s^\mathrm{FF} \). The derivation is on the page From dimensionless parameters to equipment.
Every number is computed on the server from the definitions in the section The formulas. The defaults are Example 5. Hover over or focus a symbol for its definition.
Batch elution: settings and parameters
Not updated: correct the marked field.| Quantity | Value | Unit |
|---|---|---|
| Column | ||
| Flow rate \( Q \) | – | mL/min |
| Dead time \( t_0 = \varepsilon V_\mathrm{col}/Q \) | – | min |
| Phase ratio \( F = (1-\varepsilon)/\varepsilon \) | – | – |
| Stationary-phase volume \( (1-\varepsilon) V_\mathrm{col} \) | – | mL |
| Operating point | ||
| Cut point \( s_\mathrm{cut} \) | – | – |
| Feed amount \( s^\mathrm{FF} \) | – | – |
| \( S_\mathrm{cut} = s_\mathrm{cut} + s^\mathrm{FF} \) | – | – |
| Equipment | ||
| Injection time \( \Delta t^\mathrm{FF} \) \( (= \Delta t^\mathrm{F}) \) | – | min |
| Injection volume \( V_\mathrm{inj} = Q\,\Delta t^\mathrm{FF} \) | – | mL |
| Cut time \( t_\mathrm{cut} \), from the start of the injection | – | min |
| Volume pumped up to the cut, \( Q\,t_\mathrm{cut} \) | – | mL |
Values you entered are marked input.
Four-zone SMB: flows and parameters
Not updated: correct the marked field.| Zone | \( s_j \) | \( Q_j \) mL/min |
|---|---|---|
| I | – | – |
| II | – | – |
| III | – | – |
| IV | – | – |
| Stream | \( Q \) mL/min | per switch mL |
|---|---|---|
| Eluent EL, in | – | – |
| Extract E, out | – | – |
| Fresh feed FF, in | – | – |
| Raffinate R, out | – | – |
Balance: in \( Q^\mathrm{EL} + Q^\mathrm{FF} \) = – + – = –; out \( Q^\mathrm{E} + Q^\mathrm{R} \) = – + – = – mL/min.
| Switch time \( \Delta t_\mathrm{switch} \) | – | min |
|---|---|---|
| Feed amount \( s^\mathrm{FF} = s_\mathrm{III} - s_\mathrm{II} \) | – | – |
| \( S_\mathrm{II} = s_\mathrm{II} + s^\mathrm{FF}/N_\mathrm{SMB} \), with \( N_\mathrm{SMB} = 1 \) | – | – |
| Phase ratio \( F = (1-\varepsilon)/\varepsilon \) | – | – |
| Stationary-phase volume per column \( (1-\varepsilon) V_\mathrm{col} \) | – | mL |
Values you entered are marked input. The feed volume per switch, \( Q^\mathrm{FF}\,\Delta t_\mathrm{switch} = s^\mathrm{FF}(1-\varepsilon)V_\mathrm{col} \), equals the injection volume of batch elution at the same \( s^\mathrm{FF} \).
The formulas
With the column volume \( V_\mathrm{col} \), the total porosity \( \varepsilon \), the stationary-phase volume \( (1-\varepsilon)V_\mathrm{col} \) and the phase ratio \( F = (1-\varepsilon)/\varepsilon \), the UD parameters are defined as on the Scientific basis page. They hold for any isotherm.
Batch elution
The flow rate \( Q \) is the same during the injection and the elution, and \( t_0 = \varepsilon V_\mathrm{col}/Q \).
\[ s^\mathrm{FF} = \frac{Q\,\Delta t^\mathrm{FF}}{(1-\varepsilon)V_\mathrm{col}} = \frac{\Delta t^\mathrm{FF}}{t_0}F^{-1} \]
\[ \begin{aligned} s_\mathrm{cut} &= \frac{Q\left(t_\mathrm{cut}-\Delta t^\mathrm{F}\right)-\varepsilon V_\mathrm{col}}{(1-\varepsilon)V_\mathrm{col}} \\ &= \left(\frac{t_\mathrm{cut}-\Delta t^\mathrm{F}}{t_0}-1\right)F^{-1} \end{aligned} \]
\( \Delta t^\mathrm{F} \) is the duration of the column feed and \( \Delta t^\mathrm{FF} \) that of the fresh feed. In batch elution all of the column feed is fresh feed, so \( \Delta t^\mathrm{F} = \Delta t^\mathrm{FF} \). Solved for the settings:
\[ \begin{gathered} \Delta t^\mathrm{FF} = \frac{s^\mathrm{FF}(1-\varepsilon)V_\mathrm{col}}{Q}, \qquad V_\mathrm{inj} = Q\,\Delta t^\mathrm{FF} \\ t_\mathrm{cut} = \Delta t^\mathrm{FF} + t_0\left(1 + F s_\mathrm{cut}\right) \end{gathered} \]
The cut counted from the start of the injection is \( S_\mathrm{cut} = s_\mathrm{cut} + s^\mathrm{FF} = \left(t_\mathrm{cut}/t_0 - 1\right)F^{-1} \). It is negative when the cut comes before \( t_0 \), when only the liquid that filled the column has left; the converter warns about that.
Four-zone SMB
Each zone \( j \) carries the flow \( Q_j \) during the switch time. Its parameter is that of an SMB event:
\[ s_j = \frac{Q_j\,\Delta t_\mathrm{switch}-\varepsilon V_\mathrm{col}}{(1-\varepsilon)V_\mathrm{col}} \]
for the zones \( j = \mathrm{I}, \mathrm{II}, \mathrm{III}, \mathrm{IV} \), or, solved for the flow,
\[ Q_j = \frac{s_j (1-\varepsilon)V_\mathrm{col} + \varepsilon V_\mathrm{col}}{\Delta t_\mathrm{switch}} \]
\( V_\mathrm{col} \) is the volume of one column, whatever the number of columns per zone. The balances of the four nodes give the port flows, and together they close the overall balance:
\[ \begin{aligned} Q^\mathrm{EL} &= Q_\mathrm{I}-Q_\mathrm{IV}, & Q^\mathrm{E} &= Q_\mathrm{I}-Q_\mathrm{II}, \\ Q^\mathrm{FF} &= Q_\mathrm{III}-Q_\mathrm{II}, & Q^\mathrm{R} &= Q_\mathrm{III}-Q_\mathrm{IV} \end{aligned} \]
\[ Q^\mathrm{EL} + Q^\mathrm{FF} = Q^\mathrm{E} + Q^\mathrm{R} \]
The fresh feed enters for the whole switch time, \( \Delta t^\mathrm{FF} = \Delta t_\mathrm{switch} \), and the classical SMB has one SMB event per cycle, \( N_\mathrm{SMB} = 1 \). Hence
\[ s^\mathrm{FF} = \frac{Q^\mathrm{FF}\,\Delta t_\mathrm{switch}}{(1-\varepsilon)V_\mathrm{col}} = s_\mathrm{III} - s_\mathrm{II} \]
\[ S_\mathrm{II} = s_\mathrm{II} + \frac{s^\mathrm{FF}}{N_\mathrm{SMB}} = s_\mathrm{III} \]
All flows are positive only if every \( s_j > -F^{-1} \) (zone flows) and \( s_\mathrm{I} > s_\mathrm{II} \), \( s_\mathrm{I} > s_\mathrm{IV} \), \( s_\mathrm{III} > s_\mathrm{II} \), \( s_\mathrm{III} > s_\mathrm{IV} \) (port flows). The converter warns when one of these fails. Whether the point separates is a question for the isotherm: find it with the design tool.
One equation is missing to fix the scale: either the switch time is given, or one flow rate, typically the largest zone flow that the pressure drop allows, and the switch time follows from it.
From batch to SMB, and what changes in practice
Under ideal conditions the operating point \( (s_\mathrm{cut}, s^\mathrm{FF}) \) of batch elution and the point \( (s_\mathrm{II}, s^\mathrm{FF}) \) of the four-zone SMB have the same feasible region, so the same numbers give the same purities. In a real column, band broadening is not the same in the two processes, so the transfer is approximate and needs a margin (see Limits of the ideal picture). The converter itself is exact for any column, but it does not know about pressure drop, the largest flow of your pumps or the volume of tubing and valves outside the columns, which delays the cut at the fraction collector and adds to the loop of an SMB.
Further reading
- Siitonen, J., Sainio, T., Unified design of chromatographic separation processes, Chem. Eng. Sci. 122 (2015) 436–451.
- Sainio, T., Unified Design of chromatographic processes with timed events: Separation of ternary mixtures, Chem. Eng. Sci. 152 (2016) 547–567.
- Sainio, T., Unified design of chromatographic separation processes: Partial separations and waste fractions, Chem. Eng. Sci. 234 (2021) 116457. Open access.
- Storti, G., Mazzotti, M., Morbidelli, M., Carrà, S., Robust design of binary countercurrent adsorption separation processes, AIChE J. 39 (1993) 471–492.